Comparison of CTDIw and Homogeneity Index on CTDI Phantoms
(1) Cideres Regional Public Hospital, Indonesia
(2) Gunung Jati Regional Hospital, Indonesia
(3) Bandung Institute of Technology, Indonesia
(4) Majalengka Regional Public Hospital, Indonesia
(5) Bandung Institute of Technology, Indonesia
(6) Bandung Institute of Technology, Indonesia
Abstract
The study was conducted to compare the Computed Tomography Dose Index Weighted (CTDIw) value values and homogeneity index on head and body phantoms with tube voltage variations. Two CTDI phantoms are Gammex (Sun Nuclear, Florida, United States) and IBA (IBA Dosimetry, Schwarzenbruck, Germany). The pencil ionization chamber was used for the measurement of CTDI. The measurements were carried out with a Toshiba Alexion 16 MSCT in a single rotation of axial mode with detector position in the phantom’s center, top, bottom, right, and left. Tube voltage values are 80 kVp, 100 kVp, and 120 kVp. Then, the homogeneity test of the phantom was carried out. The homogeneity value was obtained by measuring the average CT number in the image by determining the region of interest (ROI) at positions namely a, b, c, d, and e, In addition the ratio of the two phantoms was also carried out. The ratio was obtained from the difference of the CTDI100 value at the edge to the CTDI100 value at the center of the head and body phantom from Gammex and IBA. The results showed that the CTDIwof the Gammex head phantom are 26.83 mGy (80 kV), 53.32 mGy (100 kV) and 83.32 mGy (120 kV). While the CTDIw of the Gammex body phantom are 11.73 mGy (80 kV), 21.58 mGy (100 kV) and 36.45 mGy (120 kV). In comparison, CTDIw of the IBA head phantom are 27.01 mGy (80 kV), 55.33 mGy (100 kV) and 81.69 mGy (120 kV). While the CTDIw of the IBA body phantom are 11.85 mGy (80 kV), 23.32 mGy (100 kV) and 35.00 mGy (120 kV). The differences in CTDIw of the two phantoms were within (head phantom is 0.18 % – 2.01 %) and (body phantom is 0.13 % – 1.75 %). The difference below 5% with the p-value of the head phantom is 0.87 and body phantom is 0.89 (more than 0.05) indicates that the two phantoms are not significantly different because the two phantoms are made of the same material. The average ratio for the Gammex head phantom is 1.12 – 1.28, while the IBA head phantom is 1.07 – 1.28. Then the average ratio for the Gammex body phantom is 2.03 – 2.56, while for the IBA body phantom is 1.91 – 2.59 which indicates that the head phantom produces a more uniform dose distribution compared to a body phantom. The average homogeneity value of the IBA phantom is 90.52 % and the average homogeneity value of the Gammex phantom is 87.15 % (a difference of around 3.37%). This value shows that Gammex and IBA phantom have fairly good homogeneity
Keywords
Full Text:
PDFReferences
Allisy-Roberts, Penelope, et al. (2008) Farr's Physics for Medical Imaging. Saunders
Bauhs JA, Vrieze TJ, Primak AN, Bruesewitz MR, McCollough CH. (2008) CT dosimetry: comparison of measurement techniques and devices. Radiographics. 28(1):245-53. doi: 10.1148/rg.281075024. PMID: 18203941.
Bontrager, Kenneth, L. (2010) Textbook of Radiographic Positioning and Related Anatomy Seventh Edition Missouri, Mosby Inc.
CTDI Phantom, Computed Tomography Dose Index Phantom, https://www.sunnuclear.com/products/ctdi-phantoms (accessed on january 2023)
Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber T C, McCollough C, Martinoff S, Kastrati A & Schömig A. (2009) Estimated Radiation Dose Associated with Cardiac CT Angiography. J. Am. Med. Assoc. 301(5):500-7. doi: 10.1001/jama.2009.54. PMID: 19190314.
McCollough, Cynthia & Edyvean, Sue & Gould, Bob & Keat, Nicholas & Judy, Phil & Kalender, Willi & Morin, Rick & Payne, Tom & Stern, Stanley & Rothenberg, Larry. (2008). The Measurement, Reporting, and Management of Radiation Dose in CT. AAPM Report. 96.
McNitt-Gray, Michael. (2002). AAPM/RSNA physics tutorial for residents: Topics in CT - Radiation dose in CT1. Radiographics : a review publication of the Radiological Society of North America, Inc. 22. 1541-53. 10.1148/rg.226025128.
Mulkens, T., Salgado, R., Bellinck, P. (2011). Dose Optimization and Reduction in CT of the Brain and Head and Neck Region. In: Tack, D., Kalra, M., Gevenois, P. (eds) Radiation Dose from Multidetector CT. Medical Radiology(). Springer, Berlin, Heidelberg. https://doi.org/10.1007/174_2011_451
PMMA Phantoms for CT/CBCT, PMMA Phantoms for CTDI Measurement, https://www.iba-dosimetry.com/product/pmma-phantoms-for-ctcbct (accessed on januari 2023)
Söderberg M, Gunnarsson M. (2010) Automatic exposure control in computed tomography--an evaluation of systems from different manufacturers. Acta Radiol. 51(6):625-34. doi: 10.3109/02841851003698206. PMID: 20429764.
Refbacks
- There are currently no refbacks.
 This work is licensed under a Creative Commons Attribution 3.0 License. View My Stats